The EEG performance tester is compliant with the test requirements of IEC 80601-2-26:2019 and applicable for R&D, compliance and production line testing.

Overview

  • Designed for the latest EEG standard IEC 80601-2-26:2019.
  • Sophisticated electrical and shielding design enables the test system to output 1μV signal.
  • 9 channels with auto-select switches for user to test performance individually.
  • Built-in test circuits as defined in EEG standards with selection of 620kΩ/4.7nF for input impedance test.
  • Able to load and replay files in PhysioNet’s EDF format, binary format and text format.
  • IEC 80601-2-26 helper simplifies the medical standard with test sequences, options, parameters and pass criterions.
  • Software Development Kit (SDK) assists user to develop customized or automated test software with less efforts.

 

Specifications

Parameter Specification
Main output voltage accuracy  ±1% for amplitudes of 50 µVpp or higher
Main output voltage resolution (DAC resolution) 0.5µV
Frequency / pulse repetition rate accuracy ±1%
Pulse duration / timing accuracy ±1ms
Resistor tolerance ±1%
Capacitor tolerance ±5%
Precision 500:1 divider (50KΩ:100Ω ) ±0.2%
Sample rate 5kHz ± 0.1%
DC offset (fixed, noise free, from internal super capacitor) 150mV ± 1%
DC offset (variable, may include up to 50µVpp noise) Setting ±1% or ±3mV
Power supply USB +5Vdc supply (no separate power supply required) 0.5A (high power mode)
Environment 15 ~ 30°C (by design, not tested)
30 ~ 80% RH(design not tested)
Safety, EMC standards No applicable safety standards (maximum internal voltages 12Vdc) For EMC no testing performed. CE marking based on careful selection of parts, including USB protection IC, as well as special filters to reduce noise from microprocessor (8MHz) and DC/DC converter (200kHz).