The ECG simulator designed for performance testing. Suitable for compliance and production line testing. Compliant with IEC 60601-2-25, IEC 60601-2-27, IEC 60601-2-47, ANSI / AAMI EC11, ANSI / AAMI EC13, YY1079, YY1139, YY0782, YY0885, JJG 760, JJG 1041, JJG 1042.

Overview

  • Single channel performance tester for diagnostic ECG, monitoring ECG and ambulatory ECG.
  • Built-in test circuits per defined in ECG standards.
  • Excellent shielding and grounding design makes amplitude and frequency accuracy < ± 1%.
  • DC offset and pacemaker signal output can be adjustable.
  • Load ECG file function enables to load and replay recorded or programmed waveforms.
  • Offer alternative input of external generator for arbitrary waveforms.
  • Software Development Kit (SDK) assists user to develop customized or automated test software with less efforts.

 

Specifications

Parameter Specification System capability 
Main output voltage accuracy ±1% for amplitudes of 0.5mVpp or higher ±0.3%
Main output voltage resolution (DAC resolution) 2.5µV In version 4.0, a single output range is used
Frequency / pulse repetition rate accuracy ±1% ±0.1%
Pulse duration / timing accuracy (excluding pacing) ±1ms ±0.2ms
Pacing pulse width accuracy ±5µs ±1µs
Pacing pulse amplitude ±2mV pulse: ±1% ±2mV pulse: ±0.3%
accuracy, range >2mV pulse: ±10%
Range: ±2mV to ±700mV
100mV pulse: ±1% or
±5mV
Pacing pulse characteristics Rise/fall time 5µs
Overshoot <1%
Settling time <1%
 
Pacing pulse overshoot (intentional) Method A according to IEC 60601-2-27  
Resistor tolerance ±1% ±0.5%
Capacitor tolerance ±5% ±5%
Precision 1000:1 divider (100KΩ:100Ω ) ±0.2% ±0.1%
Sample rate 5kHz ± 0.1% 5kHz ± 0.05% (50ppm)
DC offset
(fixed, noise free, sourced from internal super capacitor)
300mV ± 1% 300mV ±0.1%
DC offset
(variable, may include up to 50µVpp noise)
Setting ±1% or ±3mV Setting ±1% or ±3mV