Overview
- Single channel performance tester for diagnostic ECG, monitoring ECG and ambulatory ECG.
- Built-in test circuits per defined in ECG standards.
- Excellent shielding and grounding design makes amplitude and frequency accuracy < ± 1%.
- DC offset and pacemaker signal output can be adjustable.
- Load ECG file function enables to load and replay recorded or programmed waveforms.
- Offer alternative input of external generator for arbitrary waveforms.
- Software Development Kit (SDK) assists user to develop customized or automated test software with less efforts.
Specifications
Parameter | Specification | System capability |
Main output voltage accuracy | ±1% for amplitudes of 0.5mVpp or higher | ±0.3% |
Main output voltage resolution (DAC resolution) | 2.5µV | In version 4.0, a single output range is used |
Frequency / pulse repetition rate accuracy | ±1% | ±0.1% |
Pulse duration / timing accuracy (excluding pacing) | ±1ms | ±0.2ms |
Pacing pulse width accuracy | ±5µs | ±1µs |
Pacing pulse amplitude | ±2mV pulse: ±1% | ±2mV pulse: ±0.3% |
accuracy, range | >2mV pulse: ±10% Range: ±2mV to ±700mV |
100mV pulse: ±1% or ±5mV |
Pacing pulse characteristics | Rise/fall time 5µs Overshoot <1% Settling time <1% |
|
Pacing pulse overshoot (intentional) | Method A according to IEC 60601-2-27 | |
Resistor tolerance | ±1% | ±0.5% |
Capacitor tolerance | ±5% | ±5% |
Precision 1000:1 divider (100KΩ:100Ω ) | ±0.2% | ±0.1% |
Sample rate | 5kHz ± 0.1% | 5kHz ± 0.05% (50ppm) |
DC offset (fixed, noise free, sourced from internal super capacitor) |
300mV ± 1% | 300mV ±0.1% |
DC offset (variable, may include up to 50µVpp noise) |
Setting ±1% or ±3mV | Setting ±1% or ±3mV |