Overview
- Support digital and analog data input to ensure your algorithm fully meet the standard requirements.
- AI sync function unifies original TAF into comparable TAF.
- Error data and error timing can be displayed for advanced algorithm improvement.
- Intuitive software interface allows user to complete the comparison within a few mouse clicks.
- Comparison result is identical with PhysioNet WFDB library software and can be exported to a report file.
Specification
Category | Detail |
Supported Standards | IEC 60601-2-47、ANSI / AAMI EC57、YY 0885 |
Supported Databases | MIT-BIH、AHA、NST、CU、ESC |
Supported Test Options – 17 Required Capabilities | • QRS Se、QRS +P |
• VEB Se、VEB +P、VEB FPR | |
• VEB COU Se、VEB COU +P、VEB S_Run Se、VEB S_Run +P、VEB L_Run Se、VEB L _Run +P | |
• RMS heart rate error | |
• % beats missed、% N and S missed 、% V missed、% F missed | |
、Total SHUTDOWN time | |
SupportedTest Options – 25 Optional Capabilities | • SVEB Se、SVEB +P、SVEB FPR |
• SVEB COU Se、SVEB COU +P、SVEB S_Run Se、SVEB S_Run +P、SVEB L_Run Se、SVEB L_Run +P | |
• VF EP Se、VF EP +P、VF DU Se、VF DU +P、VF FPR、VF Time | |
• AF EP Se、AF EP +P、AF DU Se、AF DU +P、AF FPR、AF Time | |
• ST EP Se、ST EP +P、ST DU Se、ST DU +P | |
TAF Format | • Binary-MIT annotation (*.atr) |
• Text-MIT Label (*.txt) | |
• Text-AHA Label (*.txt) | |
• Text-AAMI Label (*.txt) | |
• Text -AHA 2 fields (*.txt) | |
• Text -AAMI 2 fields (*.txt) | |
Supported Time Unit | • Sample Index (100Hz ~ 1500Hz) |
• Absolute Time | |
Supported Report Export Format | TXT, CSV, Excel, Word |