Phantoms for Phase Enhancement Assessment in MR Scanners

The Dual Cylinder phantom is suitable for testing uniform amplitude and linear phase incident along AIMD path. The phantom can be used to measure phase gradients between 5 rad/m and 12.5 rad/m and supports the testing of elongated structures (up to 1000 mm).

The phantom can be used in both clinical and laboratory environments (compatible with MITS1.5 or MITS3.0).
An implant holder can be provided upon request.


Material Transparent Plexiglass (PMMA)
Liquid compatibility Compatible with sugar- and oil-based tissue simulating media.
  Not compatible with Triton or DGBE-based liquids.
Dimensions Diameter 440 mm; Height 200 mm
Wall thickness Sidewall 5 ± 0.2 mm
  Bottom 10 ± 0.2 mm
Racetrack material FR-4 Glass reinforced epoxy
Racetrack dimensions Diameters: 160 mm, 180 mm, 200 mm, 265 mm, 320 mm and 400 mm
  Height: 20 mm
Empty weight Approximately 9.3 kg

  • Product list